REALITi 360 — Version Update 11.7.0
New features for ultimate performance available in REALITi 360 V. 11.7.0
New Additions
CPR Ventilation Monitor
- Added scaling on ventilation bars based on manikin type or parameter settings
- Added numeric label for max volume based on manikin type or parameter settings
- Included a line for each max volume and set at 90% of the full scale
- Enabled clipping of values greater than the full-scale
CPR Depth Monitor
- Added CPR depth configuration from Control on Zoll X Series, Zoll X Advanced, Zoll Propaq MD, Corpuls 3, Zoll R Series, Generic Defib, and CPR Console
- Enabled conversion of units for compression depths
- Improved CPR bar displays for compression depths
- Adjusted CPR bars to match CPR depth values
- Resolved the issue on the monitor with a CPR device crashing upon connection with the Control
CPR Depth Control / Facilitator
- Added CPR depth configuration in Parameter Settings
- Added conversion of units for Adequate Depth
- Added conversion of units for Maximum Depth
- Added CPR manikin type drop-down option for Automatic, Newborn, Child, and Adult
CPR Depth
- Resolved the issue where compression depth values are not updating properly when switching to another monitor
- Fixed the issue where adequate compression depth value is not displayed on the correct level
- Rectified the issue of the monitor crashing when switching to other monitors while Atlas Junior is connected
- Corrected Corpuls3 and Corpuls 3T values showing zeros instead of the default values
Improvements
Atlas
- Increased length of low RR CO2 when using SimConnect or Atlas manikin for ventilation
- Fixed Atlas CPR not affecting HR
- Added CPR depth configuration from Control on Zoll X Series, Zoll X Advanced, Zoll Propaq MD, Corpuls 3, Zoll R Series, Generic Defib, and CPR Console. Enabled conversion of units for compression depths. Improved CPR bar displays for compression depths. Adjusted CPR bars to match CPR depth values.
- Fixed AED not working from “movement detected” due to compressions state not ending
- Made the connection process for Atlas Junior the same as in 11.6.0
Corpuls 1
- Resolved the issue on Hamilton T1 where the alarm continues after quitting a scenario
- Corrected therapy socket influence on electrodes
Corpuls 3/3T
- Fixed issue with metronome count not reverting to the original configured ratio
- Fixed issue with Start Analysis being played just after powering on, no delay
- Fixed issue with metronome not resetting to Adult 30:2 when C3/C3T is powered off
- Fixed memory leak in Corpuls 3/3T which resulted in a crash when used with HamiltonT1
Defigard Touch 7
- Added parameter field flash on NIBP, Temp, ETCO2, and RR when values are off limits
- Enabled flash on bigger NIBP field
- Improved priority alarms to show in the alarm status field properly
- Corrected precedence of alarm message from any parameter over “ECG: Cable not detected” message
- Added “ECG: VF/VT” if HR alarm still persists for a certain period of time for any patient type
- Included “ECG: Asystole” message when HR is 0
- Improved alarm light to change from orange to red if alarm is for VF, VT, or Asystole
- Improved alarm light to show orange for medium and low alarms
- Resolved the issue of Pulse field not flashing when pulse is off limits
- Rectified “SpO2: Pulse off limits” message issue from not appearing when there is a pulse rate alarm
- Integrated pulse alarm limits to the alarms; the value on the pulse will now determine whether to show the alarm based on these limits
Hamilton T1
- Resolved the issue on Hamilton T1 where the alarm continues after quitting a scenario
Multiple Monitors / Screens
- Resolved the issue of the additional monitor not showing the correct state of pleth based on the primary monitor
- Fixed ECG waveform not changed if the new waveform was the same as the previously loaded scenario
Zoll R
- Resolved the issue with Filtered ECG not getting deactivated by Control