News & Updates

Return to News List

REALITi version 11.7.0 ready for deployment!

|
REALITi Version 11.7.0

New features for ultimate performance available in REALITi 360 V. 11.7.0

New Additions

CPR Ventilation Monitor

  • Added scaling on ventilation bars based on manikin type or parameter settings
  • Added numeric label for max volume based on manikin type or parameter settings
  • Included a line for each max volume and set at 90% of the full scale
  • Enabled clipping of values greater than the full-scale

CPR Depth Monitor

  • Added CPR depth configuration from Control on Zoll X Series, Zoll X Advanced, Zoll Propaq MD, Corpuls 3, Zoll R Series, Generic Defib, and CPR Console
  • Enabled conversion of units for compression depths
  • Improved CPR bar displays for compression depths
  • Adjusted CPR bars to match CPR depth values
  • Resolved the issue on the monitor with a CPR device crashing upon connection with the Control

CPR Depth Control / Facilitator

  • Added CPR depth configuration in Parameter Settings
  • Added conversion of units for Adequate Depth
  • Added conversion of units for Maximum Depth
  • Added CPR manikin type drop-down option for Automatic, Newborn, Child, and Adult

CPR Depth

  • Resolved the issue where compression depth values are not updating properly when switching to another monitor
  • Fixed the issue where adequate compression depth value is not displayed on the correct level
  • Rectified the issue of the monitor crashing when switching to other monitors while Atlas Junior is connected
  • Corrected Corpuls3 and Corpuls 3T values showing zeros instead of the default values

Improvements

Atlas

  • Increased length of low RR CO2 when using SimConnect or Atlas manikin for ventilation
  • Fixed Atlas CPR not affecting HR
  • Added CPR depth configuration from Control on Zoll X Series, Zoll X Advanced, Zoll Propaq MD, Corpuls 3, Zoll R Series, Generic Defib, and CPR Console. Enabled conversion of units for compression depths. Improved CPR bar displays for compression depths. Adjusted CPR bars to match CPR depth values.
  • Fixed AED not working from “movement detected” due to compressions state not ending
  • Made the connection process for Atlas Junior the same as in 11.6.0

Corpuls 1

  • Resolved the issue on Hamilton T1 where the alarm continues after quitting a scenario
  • Corrected therapy socket influence on electrodes

Corpuls 3/3T

  • Fixed issue with metronome count not reverting to the original configured ratio
  • Fixed issue with Start Analysis being played just after powering on, no delay
  • Fixed issue with metronome not resetting to Adult 30:2 when C3/C3T is powered off
  • Fixed memory leak in Corpuls 3/3T which resulted in a crash when used with HamiltonT1

Defigard Touch 7

  • Added parameter field flash on NIBP, Temp, ETCO2, and RR when values are off limits
  • Enabled flash on bigger NIBP field
  • Improved priority alarms to show in the alarm status field properly
  • Corrected precedence of alarm message from any parameter over “ECG: Cable not detected” message
  • Added “ECG: VF/VT” if HR alarm still persists for a certain period of time for any patient type
  • Included “ECG: Asystole” message when HR is 0
  • Improved alarm light to change from orange to red if alarm is for VF, VT, or Asystole
  • Improved alarm light to show orange for medium and low alarms
  • Resolved the issue of Pulse field not flashing when pulse is off limits
  • Rectified “SpO2: Pulse off limits” message issue from not appearing when there is a pulse rate alarm
  • Integrated pulse alarm limits to the alarms; the value on the pulse will now determine whether to show the alarm based on these limits

Hamilton T1

  • Resolved the issue on Hamilton T1 where the alarm continues after quitting a scenario

Multiple Monitors / Screens

  • Resolved the issue of the additional monitor not showing the correct state of pleth based on the primary monitor
  • Fixed ECG waveform not changed if the new waveform was the same as the previously loaded scenario

Zoll R

  • Resolved the issue with Filtered ECG not getting deactivated by Control

Visit the App Store today to update your REALITi 360 system.

Contatta iSimulate

Contattaci e scopri tutte le funzionalità dei prodotti iSimulate