REALITi Version 11.12.0 Improvements Are Here

REALITi Version 11.12.0 Improvements Are Here
Prepare to enhance your REALITi 360 experience with our latest update! Explore improvements that will enhance your customers' experience.
Latest Improvements
Atlas
- Fixed auto adjust manikin ages for the following monitors CPR feedback dashboard: Zoll X Series/Advanced/Propaq MD, Zoll R Series, Generic Defib, Schiller Defigard 7. CPR depth ranges can be customized on control
- Fixed auto adjust age changes for center compression feedback and compression depth
Corpuls
- Fixed a bug on the C3/C3T where Demand mode was possible with no electrodes attached
- Re-added 'Stimulation fortsetzen' to C3 Pacing
- Resolved the issue of 'Start Analysis Automatically' setting from C3 affecting C1 text prompts display
- Added correct text translation “PreShock CPR durchführen” for C3
- Corrected text prompts ‘Analyse starten’ and ‘Start Analysis’ for C1
- Created tutorial window for first login
- Added options to hide tutorial once or forever
- Improved Pacer pause and continue pacing prompts
- Amended Pacer pause and continue pacing DE translations
- Resolved the issue of being unable to charge after powering on with the manual button
- Resolved the issue of overlapping DE voice prompts while the “Patient nicht berühren” prompt was playing
- Resolved the issue of prompting the user to start analysis when it was not required after a shock delivery during the switch from Manual to AED
- Resolved the issue where the pie chart gets reset incorrectly during a switch from Manual to AED (and vice versa) while charging; the pie chart reset will now only occur if it is fully charged
Facilitator/Control
- Resolved the issue of ART/BP line showing zero values in Generic screen
- Added support for multiple scenario folder levels
Hamilton T1
- Added a ‘Feature not Simulated’ popup to the O2 enrichment key for Hamilton T1
- Added the day/night toggle for Hamilton T1
Lifepak 15
- Added preShock CPR user input in LIFEPAK 15 Settings
- Added preShock CPR in LifePak15
- Added preShock CPR duration restrictions: values 0-14 are considered Off state, values 15-30 activate preShock, and values above 30 are capped at 30
- Set preShock CPR duration default at 15s
- Implemented printing out of waveforms currently displayed on the screen
- Resolved the issue of ‘Shock Advised’ audio overlap
- Resolved the issue of “Analyzing now—Stand clear” text abruptly flashing when the Analyze button is pressed during a CPR countdown
Nihon Kohden - EMS1052
- Default lead sources in different modes improvement
- Added CPR assist linkage icons
- Created connectivity status for connected and disconnected CPR device
- Added temporary static CPR-Comp image for CPR device connections for AED mode only
- Fixed screen reloading when switching between modes
- Improved translations for the Nihon Kohden EMS 1052
- Resolved the issue on Lead change in 12-Lead ECG where it displayed incorrect and incomplete leads
- Resolved the issue on Lead change in 12-Lead ECG where it displayed incorrect and incomplete leads
- Improved Lead changes when switching to 12-Lead ECG where it will change source Pad to II and other leads will remain as is
- Corrected lead source when switching from Large Numerics to 12-Lead ECG
REALITi Community
- Added Reset password window for first new users
- Added Password restrictions/conditions for character input
- Added warning for mismatched passwords
Visit the App Store today to update your REALITi 360 system.